Monkey Interleukin 4 (IL-4) ELISA Kit Procedure

Monkey Interleukin 4 (IL-4) ELISA Kit Procedure!

Monkey Interleukin 4 (IL-4) steps:

1. Add 50 ul of Standards and serum samples to the wells of the corresponding reaction plates.

2. Add 100 ul of enzyme conjugate (Conjugate) to each well and mix gently for 30 seconds (important) at room temperature for 60 minutes.

3. Washing the plate: Drain the liquid in the plate, wash the reaction plate with distilled water or washing solution (ordinary washing solution, self-matching), and remove the water droplets (dry on the thick stack of absorbent paper); this is repeated 5 times.

4. Add 100 ul of coloring solution to each well and mix gently for 10 seconds at room temperature for 20 minutes.

5. Add 50 ul of Stop Solution to each well. Mix gently for 30 seconds; read OD at 450 nm in 30 minutes. Elisa kit

Sample processing and requirements:

1. Serum: The blood is naturally coagulated at room temperature for 10-20 minutes and centrifuged for about 20 minutes (2000-3000 rpm). The supernatant is carefully collected, and if precipitation occurs during storage, it should be centrifuged again.

2. Plasma: EDTA or sodium citrate should be selected as an anticoagulant according to the requirements of the specimen. After mixing for 10-20 minutes, centrifuge for about 20 minutes (2000-3000 rpm). The supernatant is carefully collected, and if a precipitate forms during storage, it should be centrifuged again.

3. Urine: Collect with a sterile tube and centrifuge for about 20 minutes (2000-3000 rpm). The supernatant is carefully collected, and if a precipitate forms during storage, it should be centrifuged again. The chest and ascites and cerebrospinal fluid are referred to.

4. Cell culture supernatant: When detecting secreted components, collect them in a sterile tube. Centrifuge for about 20 minutes (2000-3000 rpm). Collect the supernatant carefully. When the components in the cells were detected, the cell suspension was diluted with PBS (pH 7.2-7.4), and the cell concentration reached about 1 million/ml. By repeated freezing and thawing, the cells are destroyed and the intracellular components are released. Centrifuge for about 20 minutes (2000-3000 rpm). Collect the supernatant carefully. If a precipitate forms during storage, it should be centrifuged again.

5. Tissue specimen: After cutting the specimen, weigh the weight. Add a certain amount of PBS, pH 7.4. It was quickly frozen and stored in liquid nitrogen for use. The specimen still maintains a temperature of 2-8 ° C after melting. A certain amount of PBS (pH 7.4) was added, and the specimen was homogenized by hand or homogenizer. Centrifuge for about 20 minutes (2000-3000 rpm). Collect the supernatant carefully. One part of the package is to be tested, and the rest is frozen for use.

6. The specimens should be extracted as soon as possible after collection, and the extraction should be carried out according to the relevant literature. The experiment should be carried out as soon as possible after extraction. If the test cannot be performed immediately, the specimen can be stored at -20 °C, but repeated freezing and thawing should be avoided.

7. Samples containing NaN3 could not be detected because NaN3 inhibited horseradish peroxidase (HRP) activity.

Monkey Interleukin 4 (IL-4) ELISA Kit Note:

1. The kit should be taken out from the refrigerated environment and should be used at room temperature for 15-30 minutes before use. If the enzyme label is unsealed after opening, the slats should be stored in a sealed bag.

2. Concentrated washing liquid may crystallize out. When diluted, it can be heated and dissolved in a water bath. The washing will not affect the result.

3. The sampler should be used for each step, and the accuracy should be corrected frequently to avoid test errors. The loading time is controlled within 5 minutes. If the number of specimens is large, it is recommended to use a gun.

4. Please make a standard curve at the same time for each measurement and make a double hole. If the content of the substance to be tested in the specimen is too high (the OD value of the sample is larger than the OD value of the standard pore), please first dilute the sample dilution with a certain multiple (n times) and then measure it. When calculating, multiply the total dilution factor ( ×n×5).

5. The sealing film is intended for single use only to avoid cross-contamination.

6. Please keep the substrate away from light.

7. Strictly follow the instructions of the manual, the test results must be based on the microplate reader reading.

8. All samples, washings and various wastes should be treated as infectious materials.

9. The different batch components of this reagent must not be mixed.

10. If it is different from the English manual, the English manual shall prevail.

Hand Protective Gloves

Dongying Hong Xing Labor Protection Products Co.,Ltd. , https://www.hongxinglabor.com